Accuracy, reliability, linearity of AccuTrend lactate and Lactate Pro analyzers versus EBIO plus analyzer

Articolo
Data di Pubblicazione:
2009
Abstract:
Abstract The accuracy, reliability and linearity in two hand-portable lactate analyzers, the Accutrend Lactate®
(AL) and the Lactate Pro™ (LP) versus the EBIO® plus
analyzer (EP) were evaluated. For accuracy, duplicate samples
recorded on both the AL and LP revealed an overall
average diVerence versus EP (P < 0.05). The limits of
agreement between AL and EP were ¡0.7 to +1.0, and
¡1.3 to +1.5 mM between LP and EP. Reliability of AL
and LP was assessed at diVerent lactate concentrations;
coeYcient of variation ranged between 1.8 and 3.3% for
AL and between 2.8 and 5.0% for LP. AL and LP had a
good reliability for intra-, inter-analyzers, and between test
strips (ICC r = 0.999). The linearity was determined versus
the EP as reference. The slope coeYcient of AL (1.0394)
was closer to 1 than that of LP (1.1053). On these bases,
AL and LP can be individually considered suitable for the
sports research Weld.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
Portable lactate analyzers; Accuracy; Reliability; Linearity
Elenco autori:
Baldari, C; Bonavolontà, V; Emerenziani, Gp; Gallotta, Mc; Silva, Aj; Guidetti, L
Autori di Ateneo:
BALDARI CARLO
Link alla scheda completa:
https://iris.uniecampus.it/handle/11389/25918
Pubblicato in:
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY